Plastic Surgery New Jersey Cosmetic Surgery Plastic Surgery Facelift New Jersey NJ Cosmetic Surgery Plastic Surgery Breast New Jersey NJ Cosmetic Surgery Plastic Surgery Body Contouring New Jersey NJ Cosmetic Surgery Plastic Surgery Skin Tightening New Jersey NJ Cosmetic Surgery Plastic Surgery New Jersey NJ
Cosmetic Surgery Plastic Surgery New Jersey NJ Cosmetic Surgery Plastic Surgery New Jersey NJ
Plastic Surgery | Cosmetic Surgery | New Jersey
Dr. Martin Moskovitz M.D. | Plastic Surgeon | New Jersey
Cosmetic Surgeon New Jersey
Facelift | Breast Augmentation | Liposuction | New Jersey
Facelift | Breast Augmentation | Liposuction | New Jersey
Best Plastic Surgeon New Jersey
Plastic Surgery New Jersey
Breast Augmentation | Laser Liposuction | New Jersey
Dr. Martin Moskovitz MD | 140 Route 17 North | Suite 105 | Paramus New Jersey 07652 | Tel: (201) 255 - 1101
Plastic Surgery New Jersery

Image Plastic Surgery

Cosmetic Surgery Plastic Surgery Blog New Jersey NJ

Cosmetic Surgery Plastic Surgery Patient Registration New Jersey NJ

Cosmetic Surgery Plastic Surgery Photos New Jersey NJ
American Society of Plastic Surgeons Member New Jersey NJ
 
Cosmetic Breast Surgery | Reconstructive Breast Surgery | New Jersey
 

Microvascular Anastomoses Utilizing New Intravascular Stents

Martin J. Moskovitz, MD*
Lawrence Bass, MD
Ling Zhang, MD
John W. Siebert, MD

Evolving microsurgical technique has allowed for the anastomosis of smaller diameter vessels. Standard suture anastomoses cause a measure of stenosis and intimal damage during application and therefore cannot be dependably used in the anastomosis of very small vessels. We developed and tested a fibrin glue-based anastomosis applied over a meltable stent made of mono- di- and tri-glycerides. In vivo rodent studies using the 0.35 mm diameter inferior epigastric artery have shown immediate and short-term patency rates better than those associated with suture technique. The stent technique is significantly faster and easier than the suture technique. The glyceride stent method suffers from de­creased late patency due to aneurysm formation. In addition, we developed a glyceride-coated polyethylene glycol-based stent for use in lasered anastomoses. Work on both projects is ongoing.

Moskovitz MJ, Bass L, Zhang L, Siebert JW. Microvascular anastomoses utilizing new intravascular stents. Ann Plast Surg 1994;32:612-618

From the 'Department of General Surgery, Maimonides Medical Center, Brooklyn, NY, and the tlnstitute of Reconstructive Plastic Surgery, New York University Medical Center, New York.

Address correspondence to Dr Moskovitz, 180 Clinton St, Brooklyn, NY 11201.

 

The use of stents in vascular reconstruction is not new. One hundred years ago, Abbe [1] reported the experimental use of permanently placed in-traluminal glass tubes in the anastomosis of a canine femoral artery. In 1902, Carrel [2] sought to improve on Abbe's innovation when he used dissolvable intravascular stents composed of caramel candy cylinders. Although these experi­ments were interesting at the turn of the century, developing suture technology soon remanded these techniques to the pages of history.

In the latter half of this century, however, the diminishing size of anastomosed arteries rekin­dled the search for an anastomotic method that either supplanted sutures or made their use eas­ier and safer. Most of these efforts were focused on extraluminal cuffing [3—5] or everting ring-pin devices [6]. Routine use of smaller vessels in microvascular surgery and the continuing effort

to improve reconstructive outcomes have brought I microsurgery  back  to  the  intravascular  stent [7, 8] because both suture and coupling device techniques seem to be nearing their technical limitations.
We devised new stented anastomotic tech­niques for small vessels (range, 0.3-0.8 mm ex­ternal diameter). We believe these methods may offer a new avenue for microanastomoses.

Materials and Methods

This experiment began with a materials search to find a biocompatible stent material that would remain solid in the intravascular space, yet melt or dissolve at the appropriate time. Research led to a type of medicinal suppository base (massa esterinum) composed of mixtures of mono-, di-, and tri-glycerides (Witepsol H15, H175, H37; Hiils America, Piscataway, NJ]. These bases are insoluble in water and melt in the range of 33.5 to 36.5°C. Their constituent glycerides are normal elements in the blood stream; therefore, the stent is truly biocompatible.

To prepare the stent, the base material was melted in a water bath at 45°C and then drawn into glass micropipettes (Microcaps; Drummond Scientific, Broomall, PA) of varied internal diam­eters (0.195, 0.280, 0.340 mm) using capillary action. The glyceride casts were then allowed to cool at room temperature overnight. Prior to use, the stents were expelled from the micropipettes using appropriately sized wire plungers (Drum­mond Scientific).

The glyceride stents were used in conjunction with a sutureless, fibrin glue-based anastomosis. The glue was composed of two solutions [7], which were mixed on the artery at the time of anasto­mosis. The fibrin solution consisted of 0.25 g of bovine fibrinogen (ICN Biomedicals, Cleveland, OH) mixed with 2 ml of sterile water. The throm-bin solution consisted of 1,500 U bovine throm-bin (Sigma Chemicals, US), 0.25 ml of 0.5 mol/L calcium chloride (Sigma), 1,500 KIE of bovine lung-derived aprotinin (ICN Biomed), and 1.5 ml of sterile water. Both solutions were mixed in a 38°C water bath for at least 30 minutes prior to use.

In 'vivo testing of the glyceride stent involved 25 male Sprague-Dawley rats weighing between 380 and 480 g. After general intramuscular anes­thesia using a mixture of ketamine (Ketaset; Aveco, Ft Dodge, IA) (112.5 mg/ml) and Promace (acepromazine; Aveco) (2.5 mg/ml) given intra­muscularly at a dose of 0.07 ml/100 g body weight, the rats' groins were shaved and prepped. The choice of operative side was randomized. Through a unilateral groin incision, the inferior epigastric artery was exposed, and its diameter was measured. It was then dissected free and relieved of spasm with 1% topical lidocaine. Any branches were tied off with 11.0 nylon suture (Ethicon, Somerville, NJ). The artery was placed in a double clamp, transected with straight microscissors, and irrigated with heparinized saline (10 U/ml). An appropriately sized stent (Fig 1) was then placed into one end of the cut vessel (Fig 2). Care was taken to select a stent small t enough to easily slide into the vessel but large I enough to provide good vessel edge apposition.

The second vessel end was then placed over the stent (Fig 3), and the clamp arms were approxi­mated, thereby closing the vessel over the stent (Fig 4). The fibrin glue was then applied using two 1 ml tuberculin syringes with 27-gauge nee­dles (Monoject; Sherwood Medical, St Louis, MO). One drop of thrombin solution from one syringe was simultaneously applied to the vessel with one drop of fibrin solution from the other syringe. The vessel was left undisturbed for 5 minutes to allow the fibrin glue to set, after which 10 ml 45°C water was applied to the operative site, thereby melting the stent. The proximal clamp was released for a moment to allow a small amount of blood into the anastomotic site.

The blood/glyceride combination was mixed with a gentle squeezing motion on the artery with the blunt end of a forceps. The distal clamp was then released, followed by the proximal clamp, thereby washing the melted, emulsified glyceride out of the vessel (Fig 5). The skin was closed, and the anastomosis was re-examined on postopera­tive days 1, 7, 14, and 30. Assessment consisted of the standard arterial milking test, as well as high-frequency pulsed Doppler ultrasound exam­ination (VF-1; Crystal Biotech, Hopkinton, MA).

A control group of 14 rats was used. These rats were operated on and assessed in a similar fash­ion to the stented group except that instead of a fibrin-glue/glyceride stent anastomosis, a stan­dard, 11-0 nylon (Ethicon), 4- to 5-stitch suture anastomosis was performed. All operations were timed from either the point of first suture place­ment (control group) or the first grasping of the stent (study group), to the point of documented blood flow resumption with complete hemo-stasis.

Histological specimens were taken at each ex­amination interval. In addition, 1 specimen was taken at postoperative day 3. Specimens were re­moved from the rodent and immediately fixed in 10% buffered formalin for at least 72 hours. Speci­mens were then paraffin-embedded and sectioned. Staining consisted of standard hematoxylin/eosin stain, as well as Carstairs' stain for platelets and fibrin.Institution guidelines for the care and use of animals were adhered to throughout these stud­ies. Comparative statistical analyses consisted of student's t-tests (Microsoft Excel Version 4.0), with errors reported as standard deviation.

Results

As shown in Figure 6, the 1- and 2-hour patency rates using the glyceride stent method were 100% (25 of 25). At 24 hours, the rate was 91% (20 of 22). At 1 week, histological specimens demon­strated profound aneurysm formation, and pa­tency rates dropped to 74% (14 of 19). At 14 days, the aneurysms had grown larger, and patency rates decreased to 62% (10 of 16), whereas at 30 days, 23% (3 of 13) of the anastomoses were patent.

Histological examination of the glyceride-stented arteries showed the excellent vessel edge apposition afforded by this technique (Figs 7A, B; 8). A 3-day specimen (Fig 9) demonstrates aneu­rysm formation with continued vessel patency. Figure 10A, B shows a patent 30-day specimen with moderate aneurysmal dilation. A break­down of the normal medial wall architecture is seen, with disruption of the elastic lamellae and hyperplasia of the subintima. Figure 11 A, B dis­plays another 30-day specimen, with disruption of the medial architecture and the elastic lamel­lae evident in both low- and high-power views.
The patency results of the control suture group are seen in Figure 6. One hour postoperatively, all vessels were patent, whereas at 2 hours, 2 of the 14 vessels had closed (86% patency). At 24 hours, 55% patency was observed, which dropped to 45% at 7 days. Patency continued at 45% through days 14 and 30.

The average time to complete a stented/glued anastomosis was 18 minutes (±4 SD), whereas the sutured anastomosis required 28 minutes (±6 SD), which is a statistically significant differ­ence (p < 0.0001).

Discussion

Although the use of stents in vascular surgery dates to the 19th century, we document the first time they have been used on this small scale. Although fine sutures have provided the basis of microsurgical technique since its inception, su­ture technique is limited by the size of the vessel under study. The smaller the vessel, the greater the percentage stenosis and deformity caused by the sutures. Although the change in luminal diameter of a repaired human aorta may be imperceptible, each suture in the repair of a 0.5-mm vessel slightly reduces the vessel's diameter, no matter how well the anastomosis is performed.

Use of a temporary stent and a sutureless tech­nique may obviate this problem. Because no sutures are placed, the limiting factor becomes the size of the stent itself. We have shown that stents of less than 0.2 mm are easily produced. But the stent technique clearly has its difficulties, as reflected in the 30-day patency rate of 23%.

Aneurysm formation is the most likely cause of the low 30-day patency rate. This aneurysm for­mation, however, is not unique to this experi­mental method. Maxwell and colleagues [9] de­scribed a 60% aneurysm rate in standard, sutured rat femoral anastomoses. These aneurysms were smaller than those in our experiment and did not affect the patency rate. However, theirs and others' [10] histological findings of medial necro­sis, subintimal hyperplasia, and, most impor­tantly, loss of elastic lamellar structure are identical to those presented herein.

The most likely reason for the high aneurysm-related failure rate in this study is the poor healing of the media and the elastic lamellae, which is brought on by the constant pressure of the arterial system on the fibrin glue. Despite adequate strength to prevent bursting, fibrin glue lacks the durability to hold the media in close enough apposition to allow proper healing over time. Sutures, in contrast, lose a minimal amount of their strength over time and keep the media better approximated during the postoperative healing period. The fibrin glue-based anastomo­sis ultimately fails late (7-30 days) because local flow aberrations in the aneurysmal area throm-bose the vessel. Sutured anastomoses, in contrast, fail early because the technical difficulty of su­ture placement causes intimal damage and steno­sis and results in early (2-24 hr) thrombosis of the vessel.

Our laboratory is currently exploring ways to circumvent the fibrin glue weakness. One method under study is to increase the fibrin glue's strength by the simple addition of coagulation factor XIII. Factor XIII, or fibrin cross-linking factor, is found in normal plasma, but it is absent from our commercially derived preparation. (The factor was not commercially available at the time of this experiment, although it is made by Im-muno Research, NY). This omission has been cited [11] as a potentially serious flaw in fibrin glue preparation, which could yield suboptimal fibrin mesh strength. The inclusion of factor XIII may yield a better long-term patency. Likewise, transforming growth factor-/3 has been shown to increase postoperative intra-abdominal adhesion formation in rats. If the current theory on adhe­sion formation is correct and these adhesions form from incompletely degraded fibrin aggre­gates, it should be possible to intentionally in­hibit fibrin degradation and promote a more permanent fibrin clot.

Another method to decrease aneurysm forma­tion could be the use of a cuffed end-to-end anastomosis. By cuffing the anastomosis, a ring of vessel wall would surround the anastomosis and afford an extra degree of protection from the constant outward forces of the arterial system.

The stented/fibrin glue-based system is signif­icantly faster than standard suture techniques. Although a sutured anastomosis required an av­erage of 28 minutes, a stented/glued anastomosis took only 18 minutes; 5 of those minutes spent waiting for the fibrin to cure. Therefore, during this cure time, another vessel could be started, and the anastomosis of two 0.35-mm vessels could be performed in less than 25 minutes.

The glyceride stent is only one of several new techniques our laboratory is experimenting with. We are currently working on a two-staged stent for use with laser-assisted anastomoses. The center of the stent is composed of polyethylene glycol (PEG) 4000, a water-soluble, biocompati-ble material with a melting point of approxi­mately 65°C. The PEG core is covered with a water-insoluble, glyceride coating. The PEG stent is placed in an artery analogous to that described for the glyceride stent; however, laser energy is used in place of fibrin glue to secure the anasto­mosis. The laser energy melts the outer glyceride coating but leaves the inner PEG intact to support the lumen during the laser process. After laser completion, the clamps are released and the blood flow dissolves the water-soluble PEG cen­ter. Preliminary in vivo tests are ongoing.

Use of dissolvable stents in modern microsur­gery is still a technique in its infancy. We believe that with further research, these techniques and others will someday have an active part in the practice of microsurgery.

Open Discussion

Dr William Swartz (Pittsburgh): I think it is fascinating technology, but the model seems a little bit at variance with some of the published reports of suture models. Typically, a high-qual­ity suture model is patent at 90% plus at a week, not 45% as your controls would show. Do you have an explanation for that?

Dr Moskovitz: The average suture anastomosis used is the rat femoral artery, which is roughly twice the diameter of what we are talking about here. The other thing is that obviously I was performing both of these and I do not have as much experience as most of our microvascular surgeons here. But one of our microsurgeons did perform several anastomoses on these, and pre­liminary results show roughly a 70 to 80% pa­tency at 1 day, which dropped off a little bit, maybe to 60 to 70% at 1 week and then remained steady through that period. So, there is room for improvement. We think that since the flow with the fibrin glue anastomosis weakens over time, if we can get over that problem, we probably would have a better technique for these smaller vessels.

Dr Swartz: One second comment I have about anastomotic aneurysms is that it seems to be a problem in rat femoral arteries, but to my knowl­edge is not a problem in human microvascular surgery, and I think it is model-dependent.

Dr Moskovitz: Yes. Several papers by Buncke and Maxwell have noted that also and that will be a plus for this.

References

  1. Abbe R. The surgery of the hand. NY Med J 1894;14:33-39
  2. Carrel A. La technique operatoire des anastomoses vascu-laires et la transplantation des visceres. Lyon Med 1902; 98:859
  3. Obora Y, Tamaki N, Matsumoto S. Nonsuture microvas­cular anastomosis using magnet rings. Surg Neurol 1978; 9:117-119
  4. Daniel RK, Olding M. An absorbable anastomotic device for microvascular surgery: experimental studies. Plast Reconst Surg 1984;74:329-336
  5. Coleman DJ, Timmons MJ. Non-suture external cuff tech­niques for microvascular anastomosis. Br J Plast Surg 1989;42:550-555
  6. Ostrup LT, Berggren A. The Unilink instrument system for fast and safe microvascular anastomosis. Ann Plast Surg 1986;17:521-525
  7. Kamiji T, Maeda M, Matsumoto K, et al. Microvascular anastomosis using polyethylene glycol 4000 and fibrin glue. Br J Plast Surg 1989;42:54-58
  8. Cong Z, et al. Experimental study on microvascular anas­tomosis using a dissolvable stent support in the lumen. Microsurgery 1991;12:67-71
  9. Maxwell GP, Szabo Z, Buncke HI Jr. Aneurysms after microvascular anastomoses. Incidence and pathogenesis in experimental animals. Plast Reconstr Surg 1979;63:824-829
  10. Baxter TJ, Obrien BMcC, Henderson PN, Bennett RC. The histology of small vessels following microvascular repair. Br J Surg 1972;59:617-623
  11. Saltz R, Sierra D, Feldman D, et al. Experimental and clinical applications of fibrin glue. Plast Reconstr Surg 1991;88:1005-1017

 

top

 

 

Facial Procedures Breast Procedures Procedures for the body Procedures for the skin Scarless Breast Reduction Tummy Tuck Assisted Pay Program!